Endress + Hauser CA80PH-CAE22A322+AAE1F2G3MCTMT1

Process analyzer for online measurement
of orthophosphate in aqueous solutions
Colorimetric method referring to
DIN EN 1189.
Automatic calibration/cleaning,
alarm relays, two analogue outputs
optional cooling module (blue method)for
extended
reagent shelf life
::Simple upgrade to a measuring station
with Memosens sensors
::Low maintenance costs due to high
range + shelf life of the reagents
::Functionality flexible + modular
expandable
CA80PH-74J5/0
Endress Hauser CA 80PH
CA Approval: CSA C/US General Purpose
E2 Parameter; Measuring Range: PO4-P; 0,05-10 mg/l, blue method
2 Sampling Point: 2x
A Sample Transfer: Sample collector, level detection
3 Housing; Material: Stand alone housing; plastic ASA-PC, steel coated
2 Cable Entry: G
2 Power Supply: 24 VDC +/-15%
AA >Presetting Operation Language: English
E1 >Additional Analog Output: 2x 0/4…20mA
F2 >Digital Communication: Modbus TCP, Webserver+ optional outputs
G3 >Digital Sensor Input: 4x Memosens
MC >>Additional Input, Output: 4x digital; 4x digital
TM >>Communication, Software: Mathematic
T1 >>Communication, Software: Service interface CDI, external